https://doi.org/10.1051/epjap/2010100243
A method for reducing laser heating on atomic force microscope tips
Department of Physics, Fuyang Normal College, Fuyang, Anhui, 236041, P.R. China
Corresponding author: z.kaiyin@gmail.com
Received:
8
June
2010
Revised:
5
October
2010
Accepted:
14
October
2010
Published online:
23
December
2010
This paper studies numerically the temperature reached by atomic force microscope tips illuminated by femtosecond laser pulses, in which the laser focusing dependence of tip heating is investigated in detail and the laser heating effects on different tips are discussed too. The results show that tip heating can be significantly reduced if the femtosecond laser is focused tightly on the tip apex.
© EDP Sciences, 2010