https://doi.org/10.1051/epjap/2011110110
Measurement of the temperature dependence of pulse lengths in an n-type germanium detector
1
Max-Planck-Institut für Physik, München, Germany
2
Institute for the Physics and Mathematics of the Universe, Tokyo University, Tokyo, Japan
a e-mail: isa@mppmu.mpg.de
Received:
15
March
2011
Revised:
16
May
2011
Accepted:
10
July
2011
Published online:
28
September
2011
The temperature dependence of the pulse length was measured for an 18-fold segmented n-type germanium detector in the temperature range of 77–120 K. The interactions of 122 keV photons originating from a 152Eu source were selected and pulses as observed on the core and segment electrodes were studied. In both cases, the temperature dependence can be well described by a Boltzmann-like ansatz.
© EDP Sciences, 2011